Accuracy and Reproducibility of Linear and Angular Measurements in Virtual Reality: a Validation Study

AbstractThe purpose of this experimental study is to validate linear and angular measurements acquired in a virtual reality (VR) environment via a comparison with the physical measurements. The hypotheses tested are as follows: VR linear and angular measurements (1) are equivalent to the corresponding physical measurements and (2) achieve a high degree of reproducibility. Both virtual and physical measurements were performed by two raters in four different sessions. A total of 40 linear and 15 angular measurements were acquired from three physical objects (an L-block, a hand model, and a dry skull) via the use of fiducial markers on selected locations. After both intra- and inter-rater reliability were evaluated using inter-class coefficient (ICC), equivalence between virtual and physical measurements was analyzed via pairedt test and Bland-Altman plots. The accuracy of the virtual measurements was further estimated using two one-sided tests (TOST) procedure. The reproducibility of virtual measurements was evaluated via ICC as well as the repeatability coefficient. Virtual reality measurements were equivalent to physical measurements as evidenced by a pairedt test withp values of 0.413 for linear and 0.533 for angular measurements and Bland-Altman plots in all three objects. The accuracy of virtual measurements was estimated to be 0.5  mm for linear and 0.7° for angular measurements, respectively. Reproducibility in VR measurements was high as evidenced by ICC of 1.00 for l...