A new EPR dosimeter based on glutamic acid for radiation processing application.

A new EPR dosimeter based on glutamic acid for radiation processing application. Radiat Environ Biophys. 2019 Jul 08;: Authors: Beshir WB, Soliman YS, Abdel-Fattah AA, Fahim RA Abstract L-Glutamic acid (L-GA) pellets (3.8 mm × 4 mm) and powder dosimeters were studied in the dose range of 0.1-150 kGy using the electron paramagnetic resonance (EPR) technique. The EPR spectra of irradiated L-GA pellets showed an EPR signal with eight lines, and the intensity of the signal increased with an increase of absorbed dose. The results obtained in terms of the energy-absorption coefficients suggest a similar performance of the L-GA pellets as compared to alanine pellets. The value of the temperature coefficient for the L-GA pellets during irradiation was around - 0.08%/°C which is lower than that reported for alanine dosimeter, 0.14%/°C. The influence of humidity on the pellet response was found to be negligible; i.e., the increase in response was only about 2% for a relative humidity of up to 94%. The response of L-GA powder reached stability 4 h after irradiation and continued to be stable until 47 days after irradiation. In contrast, the response of the L-GA pellet dosimeter reached stability 22 h after irradiation and continued to be stable until 8 days after irradiation. For routine applications, the L-GA pellet dosimeter should be analyzed during the stable period after irradiation, to minimize the uncertainties in dose as...
Source: Radiation and Environmental Biophysics - Category: Physics Authors: Tags: Radiat Environ Biophys Source Type: research