Validation protocols for blood pressure measuring devices: the impact of the European Society of Hypertension International Protocol and the development of a Universal Standard

In the last three decades protocols for the validation of blood pressure measuring devices have been developed by the US Association for the Advancement of Medical Instrumentation, the British Hypertension Society, the German Hypertension League, the European Society of Hypertension Working Group on blood pressure Monitoring and the International Organization for Standardization. The European Society of Hypertension International Protocol required much smaller sample size than the other protocols, aiming to reduce the time, resources and cost of validation studies and thereby increase the number of validated devices. Given its specifications, the European Society of Hypertension International Protocol was adequate for ‘high- and low-accuracy’ devices, yet assessment of ‘moderate accuracy’ devices had high uncertainty with resultant high rate of device failure. Thus, devices validated using the European Society of Hypertension International Protocol should be considered to be as accurate as those validated with the previous Association for the Advancement of Medical Instrumentation or British Hypertension Society protocols. However, the European Society of Hypertension International Protocol did not allow subgroup evaluation (arm sizes, special populations, etc). The mission of the European Society of Hypertension International Protocol to promote the concept of validation has been well achieved, as almost double studies have been published using it than all the other ...
Source: Blood Pressure Monitoring - Category: Cardiology Tags: Editorial Commentaries Source Type: research