Flow Pattern of Outflow Graft is Useful for Detecting Pump Thrombosis in a Patient with Left Ventricular Assist Device.

In this study, we propose an alternative novel ramp test using the flow velocity of outflow graft (OG). A 46-year-old man underwent continuous-flow LVAD (HeartMate II, Abbott Laboratories, Lake Forest, IL, USA) implantation for advanced heart failure due to idiopathic dilated cardiomyopathy. About 2 years after implantation, he suffered from hemolysis and symptoms of heart failure, and PT was strongly suspected. The change in LVEDD was minimal with increase in pump speed (-0.06 cm/400 rotations per minute (rpm)), suggesting PT. The systolic to diastolic velocity (S/D) ratio of OG flow, which we proposed as a new indicator of PT, also showed minimal change (-0.07/400 rpm). His clinical symptoms improved with anticoagulation therapy, and the changing slope of the S/D ratio dramatically improved to -0.92/400 rpm. Although its consistency should be verified in many other cases, this novel method can be useful for detecting PT and evaluating its clinical course. PMID: 31257336 [PubMed - as supplied by publisher]
Source: International Heart Journal - Category: Cardiology Tags: Int Heart J Source Type: research