Interventricular dyssynchrony during continuous-flow left ventricular assist device support: observation using the conductance method

In this study, we evaluated this phenomenon in a normal heart model using the adult goat (n = 5, body weight 44.5 ± 2.9 kg). A centrifugal LVAD was implanted under general anesthesia. We inserted the conductance catheter into the RV and LV to obtain the pressure–volume relationship of the two ventricles simultaneously. We defined the dyssynchronous status as the sign (plus or minus) of the LV volume-change opposite to that of RV volume-change. Dyssynchronous phase of the cardiac cycle was observed in 5.6 ± 0.65% of hearts under LVAD pump-off and 25.3 ± 3.3% under LVAD full bypass, respectively (p <  0.05). To the best of our knowledge, this is the first experimental report clarifying interventricular dyssynchrony during CF-LVAD support using the conductance method. Quantification of this phenomenon under various support conditions and assessment of influences on the right ventricular functio n will be studied in future studies.
Source: Journal of Artificial Organs - Category: Transplant Surgery Source Type: research