Measurements and Monte Carlo Simulations of 241Am Activities in Three Skull Phantoms: EURADOS-USTUR Collaboration

An international intercomparison was organized by Working Group 7, Internal Dosimetry, of the European Radiation Dosimetry Group in collaboration with Working Group 6, Computational Dosimetry, for measurement and Monte Carlo simulation of 241Am in three skull phantoms. The main objectives of this combined exercise were (1) comparison of the results of counting efficiency in fixed positions over each head phantom using different germanium detector systems, (2) calculation of the activity of 241Am in the skulls, (3) comparison of Monte Carlo simulations with measurements (spectrum and counting efficiency), and (4) comparison of phantom performance. This initiative collected knowledge on equipment, detector arrangements, calibration procedures, and phantoms used around the world for in vivo monitoring of 241Am in exposed persons, as well as on the Monte Carlo skills and tools of participants. Three skull phantoms (BfS, USTUR, and CSR phantoms) were transported from Europe (10 laboratories) to North America (United States and Canada). The BfS skull was fabricated with real human bone artificially labeled with 241Am. The USTUR skull phantom was made from the US Transuranium and Uranium Registries whole-body donor (case 0102) who was contaminated due to an occupational intake of 241Am; one-half of the skull corresponds to real contaminated bone, the other half is real human bone from a noncontaminated person. Finally, the CSR phantom was fabricated as a simple hemisphere of equival...