ACCURACY AND PRECISION OF INTRAVITREAL INJECTIONS OF ANTI–VASCULAR ENDOTHELIAL GROWTH FACTOR AGENTS IN REAL LIFE: What Is Actually in the Syringe?

Purpose: To evaluate the accuracy and precision of anti–vascular endothelial growth factor volume delivery by intravitreal injections in the clinical setup. Methods: Volume output was measured in 669 intravitreal injections administered to patients, calculated from the difference in syringe weight before and after expelling the drug. Three groups were included: prefilled bevacizumab 1.0 mL syringe (Group 1, n = 432), pre-filled ranibizumab in a small-volume syringe with low dead-space plunger design (Group 2, n = 125), and aflibercept drawn and injected using a 1.0-mL syringe (Group 3, n = 112). Accuracy was analyzed by mean absolute percentage error, and precision by coefficient of variation. Results: Volume outputs in all 3 groups were significantly different from the target of 50 μL (P