Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation. J Ophthalmol. 2019;2019:1964107 Authors: Rateb M, Abdel-Radi M, Eldaly Z, Elmohamady MN, Noor El Din A Abstract Purpose: To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. Setting: Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. Methods: Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland-Altman analysis was used to assess intertonometer agreement. Results: Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significan...
Source: Journal of Ophthalmology - Category: Opthalmology Tags: J Ophthalmol Source Type: research