Longer 6-mm Diameter Stent Retrievers Are Effective for Achieving Higher First Pass Success with Fibrin-Rich Clots

First pass success (FPS) can be defined as in vitro retrieval of clot in a single pass during mechanical thrombectomy (MT) for acute large vessel occlusion (LVO). Despite advancements in MT technology, retrieval of fibrin-rich clots remains a challenge. Therefore, the effect of stent retriever length on FPS for fibrin-rich clots was investigated by using SolitaireTM 6 × 40 versus 6 × 30 mm devices with a balloon guide catheter (BGC) or distal access catheter (DAC) and sheath, in an in vitro model of anterior circulation neurovascular anatomy. Additionally, vascular safety of the SolitaireTM 6 × 40 versus 6 × 30 mm devices was evaluated in a porcine model for differences in: luminal thrombus, inflammation, endothelial coverage, fibrin deposits, smooth muscle cell loss, elastic lamina and adventitia disruption, intimal hyperplasia, and lumen reduction, at 0, 30, and 90 days post-treatmen t. In vitro overall FPS was measured as: SolitaireTM 6 × 40 (95%) and SolitaireTM 6 × 30 (67%). FPS for clot location in middle cerebral artery was: (a) BGC (6 × 40 mm: 100%; 6 × 30 mm: 100%;n = 8); (b) DAC with 088 sheath (6 × 40 mm: 83%; 6 × 30 mm: 33%;n = 12). FPS for clot location in internal carotid artery was: (a) BGC (6 × 40 mm: 100%; 6 × 30 mm: 80%;n = 11); (b) DAC with 088 sheath (6 × 40 mm: 100%; 6 × 30 mm: 67%;n = 10). Stent length had a significant effect (Fisher ’s exact test;p #x3c; 0.05) on FPS. In vivo evaluation in the porcine model showed no difference in vascu...