Radiological tissue equivalence of deformable silicone-based chemical radiation dosimeters (FlexyDos3D).

In this study we investigated its tissue-equivalence in comparison with water and Solid Water (RMI457). We found that its effective and mean atomic numbers were 40% and 20% higher and the total interaction probabilities for kV x-ray photons were larger than those of water respectively. To assess the influence of its over-response to kV photons, its HU value was measured by kV computed tomography (CT) and was found higher than all the soft-tissue substitutes. When applied for dose calculation without correction, this effect led to an 8% overestimation in electron density via HU-value mapping and 0.65% underestimation in target dose. Furthermore, depth dose curves (PDDs) and off-axis ratios (profiles) at various beam conditions as well as the dose distribution of a full-arc VMAT plan in FlexyDos3D and reference materials were simulated by Monte Carlo, where the results showed great agreement. As indicated, FlexyDos3D exhibits excellent radiological water-equivalence for clinical MV x-ray dosimetry, while its nonwater-equivalent effect for low energy x-ray dosimetry requires necessary correction. The key findings of this study provide pertinent reference for further FlexyDos3D characterization research. PMID: 31183949 [PubMed - as supplied by publisher]