The repeating dislodgement of an Amplatzer Septal Occluder device during recovery from general anesthesia in an adult undergoing transcatheter closure of atrial septal defect : a case report.

The repeating dislodgement of an Amplatzer Septal Occluder device during recovery from general anesthesia in an adult undergoing transcatheter closure of atrial septal defect : a case report. J Med Invest. 2019;66(1.2):194-198 Authors: Abe N, Go M, Nakai K, Kato M, Tanaka K Abstract Transcatheter closure with an Amplatzer Septal Occluder (ASO) has become the standard treatment for secundum atrial septal defect (ASD). However, this procedure is associated with complications, such as device dislodgement. A 52-year-old woman was admitted with exertional dyspnea. Transesophageal echocardiography showed an ASD involving a 29 mm defect. Calculated Qp/Qs was 5.6 and all the rims were ?5 mm, with the exception of the posterior rim, which was 3 mm. Transcatheter ASD closure with an ASO was performed under general anesthesia. During emergence from anesthesia, tachycardia developed and the ASO device became dislodged. Hemodynamic changes associated with the end of anesthetic administration were believed to have led to device dislodgement. In a second transcatheter ASD closure, a low dose of propofol and remifentanil was maintained during emergence from anesthesia to reduce hemodynamic changes. However, device dislodgement occurred with nonsustained ventricular tachycardia. Finally, surgical ASD closure was performed. The large defect size, high Qp/Qs, and rim deficiency may have predisposed to device dislodgement after transcatheter ASD closure...
Source: Journal of Medical Investigation - Category: General Medicine Tags: J Med Invest Source Type: research