Sensors, Vol. 19, Pages 1448: Effect of Poly-l-Lysine Polycation on the Glucose Oxidase/Ferricyanide Composite-Based Second-Generation Blood Glucose Sensors

This study adopted hydrophilic and positively charged α-poly-l-lysine (αPLL) as the entrapment matrix for the immobilization of negatively charged glucose oxidase (GOx) and ferricyanide (FIC) on SPCEs to construct a disposable second-generation glucose biosensor. The αPLL modification is shown to facilitate the redox kinetics of FIC and ferrocyanide on the SPCEs. The SPCEs coated with 0.5 mM GOx, 99.5 mM FIC, and 5 mM αPLL had better sensitivity for glucose detection due to the appreciable effect of protonated αPLL on the promotion of electron transfer between GOx and FIC. Moreover, the SPCEs coated with 0.5 mM GOx, 99.5 mM FIC, and 5 mM αPLL were packaged as blood glucose testing strips for the measurement of glucose-containing human serum samples. The glucose testing strips had good linearity from 2.8 mM to 27.5 mM and a detection limit of 2.3 mM. Moreover, the 5 mM αPLL-based glucose testing strips had good long-term stability to maintain GOx activity in aging tests at 50 °C.
Source: Sensors - Category: Biotechnology Authors: Tags: Article Source Type: research