Monte Carlo dose verification for a single-isocenter VMAT plan in multiple brain metastases
The purpose of this study was to verify the accuracy of dose calculation algorithms of a treatment planning system for a single-isocenter volumetric modulated arc therapy (VMAT) plan in multiple brain metastases, by comparing the dose distributions of treatment planning system with those of Monte Carlo (MC) simulations. We used a multitarget phantom containing 9 acrylic balls with a diameter of 15.9 mm inside a Lucy phantom measuring 17 × 17 × 17 cm3. Seven VMAT plans were created using the multitarget phantom: 1 multitarget plan (MTP) and 6 single target plans (STP).
Source: Medical Dosimetry - Category: Radiology Authors: Kento Hoshida, Fujio Araki, Takeshi Ohno, Hirofumi Tominaga, Kazuki Komatsu, Kentaro Tamura Source Type: research