Numerical investigation of the influence of a bearing/shaft structure in an axial blood pump on the potential for device thrombosis.

Numerical investigation of the influence of a bearing/shaft structure in an axial blood pump on the potential for device thrombosis. Int J Artif Organs. 2019 Jan 10;:391398818823769 Authors: Liu GM, Jin DH, Chen HB, Hou JF, Zhang Y, Sun HS, Zhou JY, Hu SS, Gui XM Abstract Adverse events caused by flow-induced thrombus formation around the bearing/shaft of an axial blood pump remain a serious problem for axial blood pumps. Moreover, excessive anticoagulation with thrombosis around the bearing potentially increases the risk of postoperative gastrointestinal bleeding. The purpose of this study is to analyze the influence of the bearing structure on the thrombosis potential of an axial blood pump. The bearing/shaft structure was embedded into an axial blood pump numerical model. The numerical simulation and analysis are focused on the low wall shear stresses, recirculation, and residence time close to the bearing region to evaluate the potential for thrombosis around the bearing. Then, the flow field near the blood pump bearing was tested via in vitro particle image velocimetry experiments to verify the numerical results. The simulation results showed that after embedding the bearing/shaft structure a recirculation zone appeared in the outlet guide vane bearing/shaft region, the residence time increased 11-fold in comparison to the pump without the bearing/shaft structure, the scalar shear stress in the shaft surface was less than 7.8 ...
Source: The International Journal of Artificial Organs - Category: Transplant Surgery Authors: Tags: Int J Artif Organs Source Type: research