Output factors of ionization chambers and solid state detectors for mobile intraoperative radiotherapy (IORT) accelerator electron beams.

CONCLUSION: The OFs results of solid state detectors PTW microDiamond and PTW Diode E as well as the ICs with small electrode spacing distance such as IBA PPC05, PTW Advanced Markus and PTW Pinpoint are in excellent agreement with OFMC . The measurements of the other detectors evaluated in this study are less accurate, thus they should be used with caution. Particularly, PTW Diode SRS and IBA NACP-02 are not suitable and their use should be avoided in relative dosimetry measurements under high dose per pulsed (DPP) electron beams. PMID: 30632271 [PubMed - as supplied by publisher]