Intraoperative Evaluation of Cochlear Implant Electrodes Using Mobile Cone-Beam Computed Tomography
Objective:
To evaluate the electrode status during cochlear implantation (CI) using mobile cone-beam CT (mCBCT).
Study Design:
Retrospective case review.
Setting:
Tertiary referral hospital.
Patients:
Fifty-seven patients (7 bilateral surgeries, 64 ears) who underwent CI and who received intraoperative mCBCT imaging.
Intervention:
CI and CBCT during surgery.
Main Outcome Measure:
Electrode location and angular insertion depth determined by intraoperative mCBCT images.
Results:
There were six cases with cochlear malformation where intraoperative mCBCT was useful to confirm electrode location. Of 58 ears with a normal cochlear morphology, perimodiolar, straight, and mid-scalar electrodes were used in 30 (cochleostomy; 14 advance off-stylet technique cases), 27 (26 round window [RW] insertion, 1 extended round window [ERW] insertion), and 1 (RW insertion) ears, respectively. Complete scala-tympani (ST) insertion was achieved in 35 ears (14 cochleostomy, 21 RW or ERW insertion). The complete ST-insertion rate was significantly higher with RW or ERW insertion than that for cochleostomy insertion (p = 0.03), although cochleostomy insertion using the advanced off-stylet technique had a similar rate to RW or ERW insertion. The angular insertion depth values (average ± standard deviation) for perimodiolar electrodes (354.4 ± 29.44 degrees) were significantly smaller than those for Flex24 (464.8 ± 43.09 degrees) and Flex28 (518.2 ± 61.91 degr...
Source: Otology and Neurotology - Category: ENT & OMF Tags: COCHLEAR IMPLANTS Source Type: research