Comparison of reconstruction and acquisition choices for quantitative T2* maps and synthetic contrasts

ConclusionsSum of squares coil-channel combination produced artifacts in phase images, but images created with adaptive combination were artifact-free. The voxel-wise median signed difference of T2* between the vendor’s adaptive channel combination and k-space reconstruction modes was 2.9 ± 0.7 ms for white matter and 4.5 ± 0.6 ms for gray matter. Relative white matter/gray matter contrast of all synthetic images and contrast-to-noise ratio of synthetic T1-weighted images were almost equal between reconstruction modes. Our results indicate that synthetic contrasts can be generated from the vendor’s DICOM data with the adaptive combination mode without affecting the quantitative T2* values or white matter/gray matter contrast.
Source: European Journal of Radiology Open - Category: Radiology Source Type: research