In-Vivo Evaluation of Flexible Mitral Annuloplasty Rings: Are They All the Same?

CONCLUSIONS: The Simulus Flex ring showed significantly more remodeling capacity than the Duran AnCore ring, but the latter ring preserved the natural dynamics of the mitral valve to a greater degree. Insights into these biomechanical differences may facilitate strategies for surgical reconstruction in degenerative mitral valve disease, especially in patients with excessive amounts of tissue and who are at risk of systolic anterior motion of the mitral valve. PMID: 30560597 [PubMed - in process]
Source: Journal of Heart Valve Disease - Category: Cardiology Tags: J Heart Valve Dis Source Type: research