Sensors, Vol. 18, Pages 4323: Research on the Temperature Characteristics of the Photoacoustic Sensor of Glucose Solution

Sensors, Vol. 18, Pages 4323: Research on the Temperature Characteristics of the Photoacoustic Sensor of Glucose Solution Sensors doi: 10.3390/s18124323 Authors: Wei Tao Zhiqian Lu Qiaozhi He Pengfei Lv Qian Wang Hui Zhao In order to weaken the influence of temperature on photoacoustic (PA) measurements and compensate PA signals with a proposed theoretical model, the relationship of PA signal amplitude with temperature, under the condition of different glucose concentrations and different light intensities, was studied in this paper. First, the theoretical model was derived from the theory of the PA effect. Then, the temperature characteristics of the PA signals were investigated, based on the analyses of the temperature-dependent Grüneisen parameter in glucose solution. Next, the concept of a PA temperature coefficient was proposed in this paper. The result of the theoretical analysis shows that this coefficient is linear to light intensity and irrelevant to the concentration of glucose solution. Furthermore, a new concept of a PA temperature coefficient of unit light intensity was proposed in this paper. This coefficient is approximately constant, with different light intensities and solution concentrations, which is similar to the thermal expansion coefficient. After calculation, the PA temperature coefficient by the unit light intensity of glucose solution is about 0.936 bar/K. Finally, relevant experiments were carried out to verify the the...
Source: Sensors - Category: Biotechnology Authors: Tags: Article Source Type: research
More News: Biotechnology