Impact of Clinically Relevant Elliptical Deformations on the Damage Patterns of Sagging and Stretched Leaflets in a Bioprosthetic Heart Valve

AbstractAfter implantation of a transcatheter bioprosthetic heart valve its original circular circumference may become distorted, which can lead to changes in leaflet coaptation and leaflets that are stretched or sagging. This may lead to early structural deterioration of the valve as seen in some explanted transcatheter heart valves. Ourin vitro study evaluates the effect of leaflet deformations seen in elliptical configurations on the damage patterns of the leaflets, with circular valve deformation as the control. Bovine pericardial tissue heart valves were subjected to accelerated wear testing under both circular (Nā€‰=ā€‰2) and elliptical (Nā€‰=ā€‰4) configurations. The elliptical configurations were created by placing the valve inside custom-made elliptical holders, which caused the leaflets to sag or stretch. The hydrodynamic performance of the valves was monitored and high resolution images were acquired to evaluate leaflet damage p atterns over time. In the elliptically deformed valves, sagging leaflets experienced more damage from wear compared to stretched leaflets; the undistorted leaflets of the circular valves experienced the least leaflet damage. Free-edge thinning and tearing were the primary modes of damage in the sagg ing leaflets. Belly region thinning was seen in the undistorted and stretched leaflets. Leaflet and fabric tears at the commissures were seen in all valve configurations. Free-edge tearing and commissure tears were the leading cause of valve hyd...
Source: Cardiovascular Engineering and Technology - Category: Cardiology Source Type: research