Evaluation of the effect of hemolysis, bilirubin, lipemy, glucosis and natrium on mean concentration of hemoglobin (XN 2000 ® Sysmex®).

Evaluation of the effect of hemolysis, bilirubin, lipemy, glucosis and natrium on mean concentration of hemoglobin (XN 2000® Sysmex®). Ann Biol Clin (Paris). 2018 Oct 01;76(5):524-530 Authors: Désidéri-Vaillant C, Oye G, Appéré G, Carassou-Beauvais P, Pérennec V, Caron S, Sapin J, Corbé H, Berna A Abstract We evaluated the effect of haemolysis, bilirubin, lipemia, hyperglycemia and natremia on mean corpuscular haemoglobin concentration (MCHC) measured by XN 2000® (Sysmex®). Our evaluation had shown than correction is necessary for haemolysis and lipemia but not for icterus. There were not possible to reproduce variation of MCHC with hyperglycemia or variation of natremia. Blood smears show red blood cells abnormalities. PMID: 30226194 [PubMed - in process]
Source: Annales de Biologie Clinique - Category: Biochemistry Authors: Tags: Ann Biol Clin (Paris) Source Type: research