PMN-PT Single Crystal Ultrasonic Transducer With Half-Concave Geometric Design for IVUS Imaging

As the key component of intravascular ultrasound (IVUS) imaging systems, traditional commercial side-looking IVUS transducers are flat and unfocused, which limits their lateral resolution. We propose a PMN-PT single crystal IVUS transducer with a half-concave geometry. This unique configuration makes it possible to conduct geometric focusing at a desired depth. To compare performances, the proposed and the traditional flat transducer with similar dimensions were fabricated. We determined that the half-concave transducer has a slightly higher center frequency (35 MHz), significantly broader −6 dB bandwidth (54%) but a higher insertion loss (−22.4 dB) compared to the flat transducer (32 MHz, 28%, and −19.3 dB, respectively). A significant enhancement of the lateral resolution was also confirmed. The experimental results are in agreement with the finite element simulation results. This preliminary investigation suggests that the half-concave geometry design is a promising approach in the development of focused IVUS transducers with broad bandwidth and high lateral resolution.
Source: IEEE Transactions on Biomedical Engineering - Category: Biomedical Engineering Source Type: research