High-speed recording of thermal load during laser trans-epithelial corneal refractive surgery using a 750 Hz ablation system

ConclusionThe ITEC software is effective in controlling the maximum temperature rise during laser ablation in the extremely challenging situation of trans-PRK involving high ablation volumes of almost 6000 nl, potentially improving the outcomes. The ITEC system limited the maximum temperature to 39.7 °C in the epithelium, and 37.7 °C in the stroma. The epithelial temperature was always higher than stromal temperature (regardless of high or low fluence irradiation). Safety limit of 40 °C found in the literature was never reached.ResumenObjetivoEvaluar el aumento de temperatura de la córnea humana durante la queratectomía transepitelial fotorrefractiva (trans-PRK) con láser excimer de 750 Hz, utilizando software para el control inteligente del efecto térmico (ITEC).MétodosEn esta serie de casos observacional, se practicó ablación Trans-PRK en 5 ojos de 3 pacientes, utilizando un perfil asférico de un sistema láser excimer de 750 Hz. Se utilizó una cámara de infrarrojos de alta resolución con 350 fotogramas por segundo para determinar la temperatura de la superficie corneal. Las imágenes se tomaron de forma secuencial, iniciándose unos pocos segundos antes de la ablación y finalizando unos pocos segundos tras la misma. Se registró la temperatura máxima de cada pixel de una imagen dada, realizándose un gráfico con respecto al tiempo.ResultadosLa temperatura basal de la superficie ocular, inmediatamente antes del comienzo de la a...