[Comparison of King Vision video laryngoscope and Macintosh laryngoscope: a prospective randomized controlled clinical trial].

CONCLUSIONS: First pass intubation success rates were similar for the King Vision and the Macintosh (96.6% vs. 94.3%, respectively, p>0.05). King Vision resulted in a longer average time to glottic view (95% CI 0.5-1.4s, p<0.001), and time to intubation (95% CI 3-4.6s, p<0.001). The difference in time to intubation was similar when unsuccessful intubation attempts were excluded (95% CI 2.8-4.4s, p<0.001). Based on the modified Mallampati class at the preoperative visit, the King Vision improved the glottic view in significantly more patients (220 patients, 56.7%) compared with the Macintosh (180 patients, 46.4%) (p<0.001). None of the patients had peripheral oxygen desaturation below 94%. Experienced anesthesiologists may obtain similar rates of first pass intubation success and airway trauma with both laryngoscopes. King Vision requires longer times to visualize the glottis and to intubate the trachea, but does not cause additional desaturation. PMID: 30005810 [PubMed - as supplied by publisher]
Source: Revista Brasileira de Anestesiologia - Category: Anesthesiology Tags: Rev Bras Anestesiol Source Type: research