Peri-procedural routines, implantation techniques, and procedure-related complications in patients undergoing implantation of subcutaneous or transvenous automatic cardioverter-defibrillators: results of the European Snapshot Survey on S-ICD Implantation (ESSS-SICDI)

AbstractThe aim of this European Heart Rhythm Association (EHRA) prospective snapshot survey is to assess peri-procedural practices, implantation techniques, and short-term procedure-related complications associated with implantation of subcutaneous implantable cardioverter-defibrillator (S-ICD) or transvenous implantable cardioverter-defibrillator (TV-ICD), across tertiary European electrophysiology centres. An internet-based electronic questionnaire concerning implantation settings, peri-procedural routines, techniques, personnel, complications, and patient outcomes was sent to the centres routinely implanting both TV-ICDs and S-ICDs. The centres were requested to prospectively include consecutive patients implanted with either TV-ICD or S-ICD during the 8-week enrolment period. Overall, 20 centres from 6 countries enrolled 429 consecutive patients. Subcutaneous implantable cardioverter-defibrillators (20%) compared with TV-ICD were implanted mainly under general anaesthesia (72% vs. 14%), in the surgical operation room settings (69% vs. 43%), with more frequent prophylactic antibiotic administration (82% vs. 91%), and post-implant defibrillation testing (85% vs. 7%, allP <  0.05). Feasibility (implantation duration of 45 min) and short-term complication rates (4%) were comparable for S-ICDs and TV-ICDs, but the spectrum of complications varied, despite different baseline characteristics of patients undergoing the S-ICD vs. TV-ICD implantation. This EHRA snapshot s...
Source: Europace - Category: Cardiology Source Type: research