Balloon assisted retrieval of broken catheter in a 10-months old baby with patent ductus arteriosus: Case report

Publication date: June 2018Source: Journal of Indian College of Cardiology, Volume 8, Issue 2Author(s): Santosh Kumar Sinha, Mahmodula Razi, Puneet Aggarwal, Lokendra Rekwal, Anupam Singh, Nishant Kumar AbhishekhAbstractPercutaneous retrieval of intravascular foreign body can be done by snares, biopsy forceps, dormia basket or tip deflecting wires. Herein, we report percutaneous retrieval a fractured multipurpose diagnostic catheter by balloon assisted technique in a 10-months old baby while doing device closure of a patent ductus arteriosus. The catheter got fractured at the junction of shaft and the tip while manipulation to right ventricular outflow tract during antegrade wiring. The fractured proximal end was trapped at the tip of venous sheath which was successfully wired with percutaneous coronary angioplasty wire and retrieved with the help of 1.5 × 10 mm and 2 × 10 mm Maverick balloon (Boston scientific; USA). This is the first ever report of balloon assisted retrieval of intravascular foreign body.
Source: Journal of Indian College of Cardiology - Category: Cardiology Source Type: research