Improved sensitivity and limit-of-detection using a receive-only coil in magnetic particle imaging.

We present characterization measurements conducted with the prototype Rx-coil installed in a preclinical MPI scanner. The gradiometric design of the Rx-coil attenuates the unwanted signal contributions arising from the excitation field, leading to a 17 dB lower background level compared to the conventional dual-purpose coil (TxRx-coil), which is crucial for detecting low amounts of MNP. Network analyzer measurements of the frequency-dependent coil sensitivity, as well as spectral analysis of recorded MPI data demonstrate an overall increase of the coil sensitivity of about +12 dB for the Rx-coil. Comparisons of the sensitivity distributions revealed no significant degradations in terms of homogeneity for the Rx-coil compared to the TxRx-coil in an imaging volume of 6x3x3 cm³. Finally, the limit of detection was determined experimentally for each coil type using a serial dilution of MNPs, resulting in values of 133 ng of iron for the conventional TxRx-coil and 20 ng for the new Rx-coil, using an acquisition time of 2 s. A linear relationship between the reconstructed signal intensities and the iron mass in the samples was observed with coefficients of determination (R2) of above 99% in the range of the limit of detection to 3 10³ng(Fe). These results open the way for improved image quality and faster acquisition time in pre-clinical MPI scanners. PMID: 29888711 [PubMed - as supplied by publisher]