An update on quality control for the PFA-100/PFA-200. Platelets. 2018 May 24;:1-6 Authors: Favaloro EJ, Bonar R Abstract Testing of platelet function comprises a crucial element of hemostasis assessment, particularly for investigations into bleeding and/or bruising. The Platelet Function Analyzer (PFA)-100 is the most utilized primary hemostasis-screening test system available, as recently remodeled/upgraded to the PFA-200. Internal quality control (IQC) and external quality assessment (EQA) (including proficiency testing) represent critical elements of ensuring test practice quality. Although true for all tests, IQC and EQA are logistically challenging for platelet function testing, inclusive of the PFA-100/200. We accordingly update our experience with novel yet feasible approaches to both IQC and EQA of PFA-100/200. Over the past 10 years, a total of 43 challenges have been tested, with most challenges designed to mimic moderate or severe primary hemostasis defects. The current report is restricted to the last four years and has also differentially assessed PFA-100 vs. PFA-200 EQA results to identify potential variance. Numerical results for closure times (CTs) and participant-supplied interpretive comments were analyzed. Reported CTs for each challenge were within limits of expectation, and good reproducibility was evidenced by repeated challenges. Coefficients of variation (CVs) for challenges, generally ranging from 15% to 25%...