Accuracy of percutaneous pedicle screw insertion in spinal fixation of traumatic thoracic and lumbar spine fractures.

Conclusion: Percutaneous pedicle screw insertion in 60 patients receiving 410 percutaneously placed pedicle screws yielded 294 ideally placed, 56 showing pedicle encroachment, 60 (14.3%, 23 cases) exhibiting varying degrees of pedicle penetration, with 2 showing new postoperative neurological deficits (severe screw misplacement). Of interest, this technique proved to be more challenging in the thoracic spine. Larger series are needed to better establish the average rate of neurological injuries associated with percutaneous thoracic/lumbar screw misplacement. PMID: 29721356 [PubMed]
Source: Surgical Neurology International - Category: Neurosurgery Tags: Surg Neurol Int Source Type: research