Ultra–High Performance Liquid Chromatography Tandem Mass Spectrometry for Cyclosporine Analysis in Human Whole Blood and Comparison With an Antibody-Conjugated Magnetic Immunoassay

Conclusions: A rapid, simple, accurate, and reliable UHPLC-MS/MS method for CsA analysis in human whole blood was developed, validated, and applied in 85 samples. On average, 21.1% overestimation was observed in ACMIA compared with that in the UHPLC-MS/MS. Further and larger studies are required to identify whether this degree of variance could be accepted by clinicians.