Influence of heart rate on coronary calcium scores: a multi-manufacturer phantom study

AbstractTo evaluate the influence of heart rate on coronary calcium scores (CCS) using a dynamic phantom on four high-end computed tomography (CT) systems from different manufacturers. Artificial coronary arteries were moved in an anthropomorphic chest phantom at linear velocities, corresponding to<  60, 60–75 and>  75 beats per minute (bpm). Data was acquired with routinely used clinical protocols for CCS on four high-end CT systems (CT1–CT4). CCS, quantified as Agatston and mass scores were compared to reference scores at<  60 bpm. Influence of heart rate was assessed for each system with the cardiac motion susceptibility (CMS) Index. At increased heart rates (>  75 bpm), Agatston scores of the low mass calcification were similar to the reference score, while Agatston scores of the medium and high mass calcification increased significantly up to 50% for all CT systems. Threefold CMS increases at>  75 bpm in comparison with<  60 bpm were shown. For medium and high mass calcifications, significant differences in CMS between CT systems were found. Heart rate substantially influences CCS for high-end CT systems of four major manufacturers, but CT systems differ in motion susceptibility. Follow-up CCS CT scans should be acquired on the same CT system and protocol, and preferably with comparable heart rates.
Source: The International Journal of Cardiovascular Imaging - Category: Radiology Source Type: research