Accuracy of intrapartum fetal blood gas analysis by scalp sampling: A retrospective cohort study

Abstract Fetal blood gas analysis (FBGA) using scalp blood is commonly used to identify serious fetal distress. However, there is a lack of data regarding its accuracy and reliability. The aim of this study was to determine the positive predictive value (PPV) and negative predictive value (NPV) of FBGA for predicting postpartum acidosis in case of nonreassuring fetal heart rate tracings (NRFHRT). To this end, we conducted a retrospective cohort study of singleton term deliveries with NRFHRT according to Fédération Internationale de Gynécologie et d’Obstétrique and Fisher cardiotocography scores undergoing FBGA in a university hospital. The PPV and NPV of FBGA regarding neonatal acidosis (defined as a pH value ≤ 7.15 in arterial or venous umbilical cord blood) and Apgar scores indicating neonatal depression (defined as a 5-min Apgar score ≤5) were evaluated. Multivariate analysis was used to determine the influence of cardiotocography variations and the time delay between FBGA and delivery on the accuracy of FBGA. We analyzed 343 deliveries with NRFHRT. In 32 (9%) of these cases, fetal acidosis was confirmed by a postpartum umbilical cord blood pH value ≤ 7.15. In 308/343 (90%) cases, FBGA identified NRFHRT as false positive (as confirmed by nonacidotic postpartum pH values) and thus avoided unnecessary interventions such as operative delivery. The overall test accuracy of FBGA was 91%. FBGA accurately predicted postpartum cord blood pH values with a margin of ±...