CT based 3D printing is superior to transesophageal echocardiography for pre-procedure planning in left atrial appendage device closure

AbstractAccurate assessment of the left atrial appendage (LAA) is important for pre-procedure planning when utilizing device closure for stroke reduction. Sizing is traditionally done with transesophageal echocardiography (TEE) but this is not always precise. Three-dimensional (3D) printing of the LAA may be more accurate. 24 patients underwent Watchman device (WD) implantation (71  ± 11 years, 42% female). All had complete 2-dimensional TEE. Fourteen also had cardiac computed tomography (CCT) with 3D printing to produce a latex model of the LAA for pre-procedure planning. Device implantation was unsuccessful in 2 cases (one with and one without a 3D model). The model cor related perfectly with implanted device size (R2 = 1; p <  0.001), while TEE-predicted size showed inferior correlation (R2 = 0.34; 95% CI 0.23–0.98, p = 0.03). Fisher’s exact test showed the model better predicted final WD size than TEE (100 vs. 60%, p = 0.02). Use of the model was associated with reduced procedure time (70 ± 20 vs. 107 ± 53 min, p = 0.03), anesthesia time (134 ± 31 vs. 182 ± 61 min, p = 0.03), and fluoroscopy time (11 ± 4 vs. 20 ± 13 min, p = 0.02). Absence of peri-device leak was also more likely when the model was used (92 vs. 56%, p = 0.04). There were trends towards reduced trans-septal puncture to catheter removal time (50 ±  20 vs. 73 ± 36 min, p = 0.07), number of device deployments (1.3â...
Source: The International Journal of Cardiovascular Imaging - Category: Radiology Source Type: research