Performance assessment of a bleeding detection algorithm for endoscopic video based on classifier fusion method and exhaustive feature selection

Publication date: February 2018 Source:Biomedical Signal Processing and Control, Volume 40 Author(s): Farah Deeba, Monzurul Islam, Francis M. Bui, Khan A. Wahid Capsule Endoscopy (CE) is a non-invasive clinical procedure that allows examination of the entire gastrointestinal tract including parts of small intestine beyond the scope of conventional endoscope. It requires computer-aided approach for the assessment of video frames to reduce diagnosis time. This paper presents a computer-assisted method based on a classifier fusion algorithm which combines two optimized Support Vector Machine (SVM) classifiers to automatically detect bleeding regions present in CE frames. The classifiers are based on RGB and HSV color spaces; the image regions are characterized on the basis of statistical features derived from the first-order histogram probability of respective color channels. A nested cross validation strategy has been adopted for the parameter tuning and feature selection to optimize the classifiers. The optimum feature sets for the best performance are evaluated after exhaustive analysis. The proposed fusion approach achieves an average accuracy of 95%, sensitivity of 94% and specificity of 95.3% for a dataset of 8872 CE frames, which is higher than that obtained from a single classifier. Comparison with the state-of-the-art algorithms exhibits that the proposed method yields superior performance for diverse dataset.
Source: Biomedical Signal Processing and Control - Category: Biomedical Science Source Type: research