Design and pharmaceutical applications of a low-flow-rate single-nozzle impactor

Publication date: 25 November 2017 Source:International Journal of Pharmaceutics, Volume 533, Issue 1 Author(s): Hui Wang, Pallavi Bhambri, James Ivey, Reinhard Vehring A new low-flow-rate (0.5L/min) single-nozzle impactor for the concentration of dilute aerosol particles with selected pharmaceutical applications is described in this paper. The impactor can be configured up to 11 stages with a wide range of cutoff diameters from 0.6μm to 21.1μm, enabling convenient sampling of inhalable drug particles from inhalation devices and drug production processes. Its unique single-nozzle design and removable impaction plate allow direct sample transfer for subsequent compositional, morphological, solid-state, and other analysis. Agreement between the measured size distribution of fluticasone propionate particles actuated from commercial pMDI Flixotide® 250 Evohaler® and reported data in the literature verified that the impactor stages have accurate cutoff diameters as designed. The multi-stage configuration of the impactor allows rapid separation of polydisperse aerosol particles into different size classes for further characterization. Overlapping of the Raman spectra of the double-component powders from the Seretide® 250 pMDI collected using two different methods demonstrated the applicability of the impactor for a representative sampling of multi-component aerosol particles for bulk composition analysis. A time-dependent and size-dependent stability study was conducted ...
Source: International Journal of Pharmaceutics - Category: Drugs & Pharmacology Source Type: research