Estimation and comparison of the radiation effective dose during coronary computed tomography angiography examinations on single-source 64-MDCT and dual-source 128-MDCT
Goal. To estimate and compare the radiation dose associated with coronary computed tomography
angiography (CCTA) examinations on two multi-detector CT scanners (MDCT), 64-MDCT and 128-MDCT, in
daily practice. Methods. Scan parameters of 90 patients undergoing retrospective
electrocardiographic gating spiral CCTA exam were recorded during a period on a single-source
64-MDCT and a dual-source 128-MDCT, and average scan parameters were derived that were used for
dosimetry. The computed tomography dose index (CTDI) with a pencil ionisation chamber and polymethyl
methacrylate body phantom with diameter of 32 cm was measured on both scanners. The dose –length
product (DLP) was calculated and the DLP to effective dose conversion factor (for chest scan at 120
kV of 0.014 mSv mGy –1 cm –1 ) was used to estimate effective dose (ED). Results. Patients’ heart
rate, scan length, pitch factor, CTDIv, DLP and ED for 128-MDCT were 64 (5) (beats min –1...
Source: Journal of Radiological Protection - Category: Physics Authors: Daryoush Khoramian and Soroush Sistani Source Type: research