Multi-centre audit of VMAT planning and pre-treatment verification.

CONCLUSIONS: All plans fulfilled dosimetric constraints, but plan quality and complexity did not correlate and were strongly dependent on the planner and the vendor. In-house measurements cannot completely replace external audits for credentialing. PMID: 28687395 [PubMed - as supplied by publisher]
Source: Radiotherapy and Oncology : journal of the European Society for Therapeutic Radiology and Oncology - Category: Radiology Authors: Tags: Radiother Oncol Source Type: research