Comparison of DVH-based plan verification methods for VMAT: ArcCHECK-3DVH system and dynalog-based dose reconstruction.

Comparison of DVH-based plan verification methods for VMAT: ArcCHECK-3DVH system and dynalog-based dose reconstruction. J Appl Clin Med Phys. 2017 Jun 26;: Authors: Saito M, Kadoya N, Sato K, Ito K, Dobashi S, Takeda K, Onishi H, Jingu K Abstract The purpose of this study was comparing dose-volume histogram (DVH)-based plan verification methods for volumetric modulated arc therapy (VMAT) pretreatment QA. We evaluated two 3D dose reconstruction systems: ArcCHECK-3DVH system (Sun Nuclear corp.) and Varian dynalog-based dose reconstruction (DBDR) system, developed in-house. Fifteen prostate cancer patients (67.6 Gy/26 Fr), four head and neck cancer patient (66 Gy/33 Fr), and four esophagus cancer patients (60 Gy/30 Fr) treated with VMAT were studied. First, ArcCHECK measurement was performed on all plans; simultaneously, the Varian dynalog data sets that contained the actual delivered parameters (leaf positions, gantry angles, and cumulative MUs) were acquired from the Linac control system. Thereafter, the delivered 3D patient dose was reconstructed by 3DVH software (two different calculating modes were used: High Sensitivity (3DVH-HS) and Normal Sensitivity (3DVH-NS)) and in-house DBDR system. We evaluated the differences between the TPS-calculated dose and the reconstructed dose using 3D gamma passing rates and DVH dose index analysis. The average 3D gamma passing rates (3%/3 mm) between the TPS-calculated dose and the reconstruct...