Technical Note: Monte Carlo calculations of the AAPM TG-43 brachytherapy dosimetry parameters for a new titanium-encapsulated Yb-169 source.

In this study, the original MC source model was modified to accurately match the specifications of the manufactured Yb-169 source. The modified MC model was then used to obtain a complete set of the AAPM TG-43 parameters for the new titanium-encapsulated Yb-169 source. The MC-calculated dose rate constant for this titanium-encapsulated Yb-169 source was 1.19 ± 0.03 cGy·h-1·U-1, indicating no significant change from the values reported for stainless steel-encapsulated Yb-169 sources. The source anisotropy and radial dose function for the new source were also found similar to those reported for the stainless steel-encapsulated Yb-169 sources. The current results suggest that the use of titanium, instead of stainless steel, to encapsulate the Yb-169 core would not lead to any major change in the dosimetric characteristics of the Yb-169 source. The results also show that the titanium encapsulation of the Yb-169 source could be accomplished while meeting the design goals as described in the current investigators' published MC optimization study for GNRT applications. PMID: 28585277 [PubMed - as supplied by publisher]
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research