Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases.

This study therefore investigated the use of radiochromic film measurements to optimize the DLG specifically for the purpose of producing accurate VMAT plans using a flattening-filter-free (FFF) beam, for use in treating vertebral targets using a stereotactic (SABR, also known as SBRT) fractionation schedule. Four test treatments were planned using a VMAT technique, to deliver a prescription of 24 Gy in 3 fractions to four different spine SABR treatment sites. Measurements of the doses delivered by these treatments were acquired using an ionization chamber and radiographic film. These measurements were compared with the doses calculated by the treatment planning system using a range of DLG values, including a DLG identified using the conventional sliding-slit method (1.1 mm). An optimal DLG value was identified, as the value that produced the closest agreement between the planned and measured doses (1.9 mm). The accuracy of the dose calculations produced using the optimized DLG value was verified using additional radiochromic film measurements in a heterogeneous phantom. This study provided a specific initial DLG (1.9 mm) as well as a film-based optimization method, which may be used by radiotherapy centers when attempting to commission or improve an FFF VMAT-based SABR treatment programme. PMID: 28574219 [PubMed - as supplied by publisher]
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research
More News: Physics | Radiography | Study