A method to stabilize the incident X-ray energy for anomalous diffraction measurements

A method to calibrate and stabilize the incident X-ray energy for anomalous diffraction data collection is provided and has been successfully used at the single-crystal diffraction beamline 1W2B at the Beijing Synchrotron Radiation Facilities. Employing a feedback loop to control the movement of the double-crystal monochromator, this new method enables the incident X-ray energy to be kept within a 0.2   eV range at the inflection point of the absorption edge.
Source: Journal of Synchrotron Radiation - Category: Physics Authors: Tags: energy stability monochromator anomalous diffraction feedback absorption edge. research papers Source Type: research
More News: Physics