Is modification of the VVI backup mode in implantable cardioverter-defibrillators from St Jude medical required due to increased risk of inappropriate shocks?

Inappropriate implantable cardioverter-defibrillator (ICD) shock therapy is painful, stressful, and typically occurs unexpected in conscious patients and may be related to a less favourable prognosis. In our institution, we have observed four cases of multiple inappropriate ICD shocks during reset to VVI backup mode. All four patients were implanted with a St Jude Medical ICD since 2010. The reset to VVI backup mode happens as a ‘safety’ response when the ICD encounters errors in the software or hardware often due to electromagnetic interference. The ICD then operates in a simple mode, with only a ventricular fibrillation (VF) zone starting at 146 b.p.m., with shock therapy only and changes in sensitivity settings makin g the ICD more sensitive. In all cases, the reason for the multiple inappropriate shocks was that the VF zone was reached due to exercise-induced sinus tachycardia or due to oversensing during sinus rhythm. The VVI backup mode has to balance between protection from failure of ICD therapy during life -threatening ventricular arrhythmias and from inappropriate shocks. It seems the non-programmable parameters in VVI backup mode of St Jude Medical ICDs carry an unacceptable high risk of inappropriate shocks during normal rhythm as illustrated by our four cases. A higher VF zone comparable with the zones chosen by the other manufacturer would give a better balance, since it is very unlikely that a patient will need shock therapy urgently for slow ventricular ta...
Source: Europace - Category: Cardiology Source Type: research