Evaluation of a rapid antigen test for detection of Streptococcus pneumoniae in cerebrospinal fluid

This study evaluated the diagnostic accuracy of the Uni-Gold ICT to detect pneumococcal antigen in CSF specimens, compared with gold standard bacteriology and quantitative real-time PCR (qPCR) testing. CSF specimens (n=69) from patients with suspected bacterial meningitis were included in the study. 13/69 (19%) were positive and 56/69 (81%) were negative for pneumococcus by the gold standard tests. The ICT had sensitivity of 85% (55%–98%), specificity of 96% (88%–100%), positive likelihood ratio of 23.7 (6–94) and negative likelihood ratio of 0.16 (0.04–0.57). Overall, a strong correlation between the ICT and qPCR results was seen (=0.81). In contrast, CSF microscopy and culture were exceptionally insensitive. The ICT method is sufficiently robust and accurate for use in algorithms to diagnose bacterial meningitis.
Source: Journal of Clinical Pathology - Category: Pathology Authors: Tags: Meningitis, Infection (neurology), Clinical diagnostic tests Short report Source Type: research