Individual volume-based 3D gamma indices for pretreatment VMAT QA.

Individual volume-based 3D gamma indices for pretreatment VMAT QA. J Appl Clin Med Phys. 2017 Mar 20;: Authors: Yi J, Han C, Zheng X, Zhou Y, Deng Z, Xie C, Jin X, Jin F Abstract Although gamma analysis is still a widely accepted quantitative tool to analyze and report patient-specific QA for intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT), the correlation between the 2D percentage gamma passing rate (%GP), and the clinical dosimetric difference for IMRT and VMAT has been questioned. The purpose of this study was to investigate the feasibility of individual volume-based 3D gamma indices for pretreatment VMAT QA. Percentage dosimetric errors (%DE) of dose-volume histogram metrics (includes target volumes and organ at risks) between the treatment planning system and QA-reconstructed dose distribution, %GPs for individual volume and global gamma indices, as well their correlations and sensitivities were investigated for one- and two-arc VMAT plans. The %GPs of individual volumes had a higher percent of correlation with individual 15 %DE metrics compared with global %GPs. For two-arc VMAT at 2%/2 mm, 3%/3 mm, and 4%/4 mm criteria, individual volume %GPs were correlated with 9, 12, and 9 out of 15 %DE metrics, while global %GPs were correlated with only 2 out of 15 %DE metrics, respectively. For one-arc VMAT at 2%/2 mm, 3%/3 mm, and 4%/4 mm criteria, individual volume %GPs were correlated with 18, 16, a...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research