Prosthetic Aortic Valve Fixation Study: 48 Replacement Valves Analyzed Using Digital Pressure Mapping

Objective: Prostheses attachment is critical in aortic valve replacement surgery, yet reliable prosthetic security remains a challenge. Accurate techniques to analyze prosthetic fixation pressures may enable the use of fewer sutures while reducing the risk of paravalvular leaks (PVL). Methods: Customized digital thin film pressure transducers were sutured between aortic annulus models and 21-mm bioprosthetic valves with 15 × 4-mm, 12 × 4-mm, or 9 × 6-mm-wide pledgeted mattress sutures. Simulating open and minimally invasive access, 4 surgeons, blinded to data acquisition, each secured 12 valves using manual knot-tying (hand-tied [HT] or knot-pusher [KP]) or automated titanium fasteners (TFs). Real-time pressure measurements and times were recorded. Two-dimensional (2D) and 3D pressure maps were generated for all valves. Pressures less than 80 mm Hg were considered at risk for PVL. Results: Pressures under each knot (intrasuture) fell less than 80 mm Hg for 12 of 144 manual knots (5/144 HT, 7/144 KP) versus 0 of 288 TF (P