Characterization of a 2.5 MV inline portal imaging beam.

Characterization of a 2.5 MV inline portal imaging beam. J Appl Clin Med Phys. 2016;17(5):6323 Authors: Grafe JL, Owen J, Villarreal-Barajas JE, Khan RF Abstract A new megavoltage (MV) energy was recently introduced on Varian TrueBeam linear accelerators for imaging applications. This work describes the experimental characterization of a 2.5 MV inline portal imaging beam for commissioning, routine clinical use, and quality assurance purposes. The beam quality of the 2.5 MV beam was determined by measuring a percent depth dose, PDD, in water phantom for 10 × 10 cm2 field at source-to-surface distance 100 cm with a CC13 ion chamber, plane parallel Markus chamber, and GafChromic EBT3 film. Absolute dosimetric output calibration of the beam was performed using a traceable calibrated ionization chamber, following the AAPM Task Group 51 procedure. EBT3 film measurements were also performed to measure entrance dose. The output stability of the imaging beam was monitored for five months. Coincidence of 2.5 MV imaging beam with 6 MV therapy beam was verified with hidden-target cubic phantom. Image quality was studied using the Leeds and QC3 phantom. The depth of maximum dose, dmax, and percent dose at 10 cm depth were, respectively, 5.7 mm and 51.7% for CC13, 6.1 mm and 51.9% for Markus chamber, and 5.1 mm and 51.9% for EBT3 film. The 2.5 MV beam quality is slightly inferior to that of a 60Co teletherapy beam; however, an estimated kQ of 1.0...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research