Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field.

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field. J Appl Clin Med Phys. 2016;17(5):6376 Authors: Miura H, Ozawa S, Hayata M, Tsuda S, Yamada K, Nagata Y Abstract We proposed a simple visual method for evaluating the dynamic tumor tracking (DTT) accuracy of a gimbal mechanism using a light field. A single photon beam was set with a field size of 30 × 30 mm2 at a gantry angle of 90°. The center of a cube phantom was set up at the isocenter of a motion table, and 4D modeling was performed based on the tumor and infrared (IR) marker motion. After 4D modeling, the cube phantom was replaced with a sheet of paper, which was placed perpen-dicularly, and a light field was projected on the sheet of paper. The light field was recorded using a web camera in a treatment room that was as dark as possible. Calculated images from each image obtained using the camera were summed to compose a total summation image. Sinusoidal motion sequences were produced by moving the phantom with a fixed amplitude of 20 mm and different breathing periods of 2, 4, 6, and 8 s. The light field was projected on the sheet of paper under three conditions: with the moving phantom and DTT based on the motion of the phantom, with the moving phantom and non-DTT, and with a stationary phantom for comparison. The values of tracking errors using the light field were 1.12 ± 0.72, 0.31 ± 0.19, 0.27 ± 0.12, and 0.15 ...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research