A methodology for on-board CBCT imaging dose using optically stimulated luminescence detectors.

This study pre-sents a complete procedure for the CBCT dosimetry using the InLight optically-stimulated-luminescence (OSL) nanoDots. We report five dose parameters: the mean slice dose (DMSD); the cone beam dose index (CBDIW); the mean volume dose (DMVD); point-dose profile, D(FOV); and the off-field Dose. In addition, CBCT skin doses for seven pelvic tumor patients are reported. CBCT-dose mea-surement was performed on a custom-made cylindrical acrylic body phantom (50cm length, 32cm diameter). We machined 25 circular disks (2 cm thick) with grooves and holes to hold OSL-nanoDots. OSLs that showed similar sensitivities were selected and calibrated against a Farmer-type ionization-chamber (0.6 CT) before being inserted into the grooves and holes. For the phantom scan, a standard CBCT-imaging protocol (pelvic sites: 125 kVp, 80 mA and 25 ms) was used. Five dose parameters were quantified: DMSD, CBDIW, DMVD, D(FOV), and the off-field dose. The DMSD for the central slice was 31.1 ± 0.85 mGy, and CBDIW was 34.5± 0.6 mGy at 16cm FOV. The DMVD was 25.6 ± 1.1 mGy. The off-field dose was 10.5 mGy. For patients, the anterior and lateral skin doses attributable to CBCT imaging were 39.04 ± 4.4 and 27.1 ± 1.3 mGy, respectively.OSL nanoDots were convenient to use in measuring CBCT dose. The method of selecting the nanoDots greatly reduced uncertainty in the OSL measurements. Our detailed calibration procedure and CBCT dose measurements and calculations could prove useful in developin...