A cross-sectional study of the radiation dose and image quality of X-ray equipment used in IVR.

This study was conducted in 2014 at 13 medical facilities using 18 IVR X-ray systems. We evaluated image quality and simultaneously measured the radiation dose. Entrance surface doses for fluoroscopy (duration, 1 min) and cineradiography (duration, 10 s) are measured using a 20-cm-thick acrylic plate and skin dose monitor. The image quality (such as spatial resolution and low-contrast detectability) of both fluoroscopy and cineradiography was evaluated using a QC phantom. For fluoroscopy, the average entrance surface dose using the 20-cm-thick acrylic plate was 13.9 (range 2.1-28.2) mGy/min. For cineradiography, the average entrance surface dose was 24.6 (range 5.1-49.3) mGy/10 s. We found positive correlations between radiation doses and image quality scores, in general, especially for fluoroscopy. The differences in surface dose among the 18 IVR X-ray systems were high (max/min, 9.7-fold for cineradiography; 13.4-fold for fluoroscopy). The differences in image quality scores (spatial resolution, low-contrast detectability, and dynamic range) were also very large. In general, there tended to be a correlation between radiation dose and image quality. Periodical measurements of the radiation dose and image quality of the X-ray equipment used for cineradiography and fluoroscopy in IVR are necessary. The need to minimize patient exposure requires that the dose be reduced to the minimum level that will generate an image with an acceptable degree of noise. PMID: 27455503 ...