Predicting deliverability of volumetric-modulated arc therapy (VMAT) plans using aperture complexity analysis.

Predicting deliverability of volumetric-modulated arc therapy (VMAT) plans using aperture complexity analysis. J Appl Clin Med Phys. 2016;17(4):6241 Authors: Younge KC, Roberts D, Janes LA, Anderson C, Moran JM, Matuszak MM Abstract The purpose of this study was to evaluate the ability of an aperture complexity metric for volumetric-modulated arc therapy (VMAT) plans to predict plan delivery accuracy. We developed a complexity analysis tool as a plug-in script to Varian's Eclipse treatment planning system. This script reports the modulation of plans, arcs, and individual control points for VMAT plans using a previously developed complexity metric. The calculated complexities are compared to that of 649 VMAT plans previously treated at our institution from 2013 to mid-2015. We used the VMAT quality assurance (QA) results from the 649 treated plans, plus 62 plans that failed pretreatment QA, to validate the ability of the complexity metric to predict plan deliverability. We used a receiver operating characteristic (ROC) analysis to determine an appropriate complexity threshold value above which a plan should be considered for reoptimization before it moves further through our planning workflow. The average complexity metric for the 649 treated plans analyzed with the script was 0.132 mm-1 with a standard deviation of 0.036 mm-1. We found that when using a threshold complexity value of 0.180 mm-1, the true positive rate for correctly id...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research
More News: Physics | Study